Getxent tubes can be analyzed on analytical equipments. The 2 main machines are:
- Gas chromatography coupled with mass spectrometry (GC-MS)
- Liquid chromatography coupled with mass spectrometry (LC-MS)
GC-MS
Tubes can be directly analyzed by GC-MS:
- SPME
- Headspace
An example of SPME method is available below.
LC-MS
LC-MS analysis of Getxent tubes requires an extraction of the content of the Getxent tubes with solvents. We recommend the following solvents at room temperature:
- Methanol
- Dimethylformamide
- Acetone
- Isooctane
- 10% sulfuric acid solution
- 10% caustic soda solution
Splitless GC-MS method
INSTRUMENT CONTROL PARAMETERS: Â Â GC-MSD
Control Information
Sample Inlet : Â GC
Injection Source : Â Manual
Mass Spectrometer : Â Enabled
No Sample Prep method has been assigned to this method.
=========
6890 GC METHOD
=========
OVEN
- Initial temp:  40°C (On)
- Maximum temp:  260°C
- Initial time: 1.00 min
- Equilibration time: Â 3.00 min
- Ramps:
     # | Rate | Final temp | Final time
     1   10.00  200        5.00
     2   0.0(Off)
- Post temp:  0°C
- Post time: Â 0.00 min
- Run time: Â 22.00 min
FRONT INLET (SPLIT/SPLITLESS) Â Â Â Â Â BACK INLET (UNKNOWN)
Mode:Â Â Â Â Â Â Splitless
Initial temp:  250 °C (On)
Pressure:Â Â Â 16.00 psi (On)
Purge flow:Â Â 70.0 mL/min
Purge time:Â Â 3.00 min
Total flow:Â Â Â 73.8 mL/min
Gas saver:Â Â Off
Gas type:Â Â Â Helium
                      COLUMN 1                 COLUMN 2
                    Capillary Column          (not installed)
Model Number:Â Â Â Â Â Â Â SPME 1227062
                     DB WAX
Max temperature:     260 °C
Nominal length:Â Â Â Â Â Â Â 60.0 m
Nominal diameter:     250.00 µm
Nominal film thickness: 0.25 µm
Mode:Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â constant flow
Initial flow:Â Â Â Â Â Â Â Â Â Â Â Â 1.0 mL/min
Nominal init pressure:Â Â 16.01 psi
Average velocity:Â Â Â Â Â Â 26 cm/sec
Inlet:Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Front Inlet
Outlet:Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â MSD
Outlet pressure:Â Â Â Â Â Â Â vacuum
FRONT DETECTOR (NO DET) Â Â Â Â Â Â Â Â BACK DETECTOR (NO DET)
SIGNAL 1Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â SIGNAL 2
  Data rate:  20 Hz          Data rate:  20 Hz
  Type:  test plot            Type:  test plot
  Save Data:  Off            Save Data:  Off
  Zero:  0.0 (Off)            Zero:  0.0 (Off)
  Range:  0                Range:  0
  Fast Peaks:  Off           Fast Peaks:  Off
  Attenuation:  0            Attenuation:  0
COLUMN COMP 1 Â Â Â Â Â Â Â Â Â Â Â Â Â COLUMN COMP 2
(No Detectors Installed)Â Â Â Â Â Â Â Â (No Detectors Installed)
THERMAL AUX 2
- Use: Â MSD Transfer Line Heater
- Description:
- Initial temp:  250 °C (On)
- Initial time: Â 0.00 min
     # | Rate | Final temp | Final time
     1   0.0(Off)
POST RUN
- Post Time: 0.00 min
GC Injector
- Front Injector: No parameters specified
- Back Injector: No parameters specified
- Column 1 Inventory Number : AB002
- Column 2 Inventory Number :Â Â
MS ACQUISITION PARAMETERS
General Information
- Tune File : atune.u
- Acquistion Mode : Scan
MS Information
- Solvent Delay : 3.00 min
- EMV Mode : Gain Factor
- Gain Factor : 10.00
- Resulting EM Voltage : 1341
[Scan Parameters]
- Low Mass : 35.0
- High Mass : 500.0
- Threshold : 150
- Sample # : 2 A/D Samples 4
- Plot 2 low mass : 50.0
- Plot 2 high mass : 550.0
[MSZones]
- MS Source : 230 °C  maximum 250 °C
- MS Quad : 150 °C  maximum 200 °C
END OF MS ACQUISITION PARAMETERS
-----------------------------
Trace Ion Detection is OFF.
- Â EMISSION :Â Â Â Â Â 34.610
- Â ENERGYÂ Â :Â Â Â Â Â 69.922
- Â REPELLER :Â Â Â Â Â 32.468
- Â IONFOCUS :Â Â Â Â Â 88.259
- Â ENTRANCE_LE :Â Â Â Â Â 14.500
- Â EMVOLTSÂ Â Â Â :Â Â Â Â 952.941
- Actual EMVÂ :Â Â Â 952.941
- GAIN FACTOR :Â Â Â Â Â Â 0.59
- Â AMUGAINÂ Â Â Â :Â Â Â 2030.000
- Â AMUOFFSETÂ Â :Â Â Â Â 128.000
- Â FILAMENTÂ Â Â :Â Â Â Â Â Â 1.000
- Â DCPOLARITYÂ :Â Â Â Â Â Â 0.000
- Â ENTLENSOFFS :Â Â Â Â Â 19.075
- Â MASSGAINÂ Â Â :Â Â Â Â -76.000Â Â
- Â MASSOFFSETÂ :Â Â Â Â Â -8.000Â Â
END OF TUNE PARAMETERS
------------------------------------
END OF INSTRUMENT CONTROL PARAMETERS
------------------------------------
ROBOT PARAMETERS
- Cycle: SPME
- Syringe: Fiber
- Pre. Inc. Time: 00:05:00
- Incubat. Temp.: 40.0°C
- Agi. Speed: 250rpm
- Agi. On Time: 2s
- Agi. Off Time: 4s
- Vial. Penetr.: 28.0mm
- Extract Time: 00:25:00
- Desorb. to: Front
- Inj. Penetr.: 54.0mm
- Desorb. Time: 00:02:00
- Fiber Bakeout: 00:00:00
- GC-Runtime: 00:29:00